Ideal for applications requiring high counting in high-speed scintillation detectors.
The APG1500 is a charge-sensitive preamplifier characterized by a fast response time with a rise time of 20 ns or less. It is ideal for applications that require high counting rates, such as those using LaBr3(Ce) scintillation detectors.
Inquire About This Product
basic information
**Main Specifications** - Rise time < 20ns - Noise characteristic < 100μVrms - Input capacitance 220pF, 1000pF, 2200pF (switchable via rotary switch, delivered with 220pF) - Time constant < 100μs - Output: LEMO 00.250 compatible connector, output impedance 50Ω - Input: LEMO 00.250 compatible connector - Power supply: D-sub 9-pin connector (NIM standard) ±12V, below 30mA - Dimensions: 75(W) × 52.6(D) × 20(H) *excluding connectors, screws, and rubber feet - Weight: 138g *For more details, please download the PDF or feel free to contact us.*
Price range
P3
Delivery Time
※Three months after the order is received.
Applications/Examples of results
For more details, please download the PDF or feel free to contact us.
catalog(1)
Download All CatalogsCompany information
Techno AP Co., Ltd. is a company located in Hitachinaka City, Ibaraki Prefecture, that engages in the design, development, and sales of radiation measurement devices. Additionally, we also develop, manufacture, and sell electronic application devices, design and produce electronic circuits, design and produce software, design and produce measurement and control programs, design and produce automatic measurement systems using computers, and design and produce sequencer control systems.