[Technical Data] EMI Test System MR2300/RF Shielded Room MY5410
■EMI Test System MR2300
Ideal for preliminary compliance testing of official EMI tests, known as "Pre-compliance."
This is an integrated system that combines our spectrum analyzer technology, anechoic chamber technology, and antenna technology.
■Anechoic Chamber MY5410
An anechoic chamber optimal for large-scale EMI testing. It comes standard with a turntable with a diameter of 756mm and a load capacity of 100kg, designed for large EUTs.
*For more details, please refer to the PDF materials or feel free to contact us.
The MR2300/MY5410 has been developed for pre-compliance use. This enables the measurement of harmonics from switching power supplies in the 40-100MHz range, which was previously difficult. Below are the measurement results obtained with the MR2300/MY5410, compared with measurements conducted in a 10m anechoic chamber equipped with facilities equivalent to the official site, using the 3m method.