Oxygen Monitor OXYMAN Series - Inspection and Calibration

1. Daily Inspections by Users
1) Operation confirmation at the start of each work shift (displaying around 20.9% and stable)
If the displayed value is unstable, it indicates sensor degradation, and sensor replacement is necessary. *
2) Atmospheric calibration using the "CAL" button approximately once a month (displaying 20.9%)
Even within a month, if the displayed value deviates below 20.4 or above 21.4, please perform this atmospheric calibration.
This operation updates the sensor's remaining capacity.
When it reaches 1 block remaining, please promptly arrange for sensor replacement. *
* Please contact us at the contact information below.
2. Inspections and Calibration by the Manufacturer
1) For purposes other than oxygen deficiency monitoring, the contents of 1. above apply.
2) In cases where the oxygen monitor's malfunction is treated as an accident when used continuously for oxygen deficiency monitoring, in addition to 1. above, we recommend planned regular maintenance (every year).
If there are stricter regulations regarding the maintenance of safety devices at your company, please follow those.