[Patent Granted] Impact-type Accelerometer Calibration and Testing System CS18-MS
SPEKTRA Corporation CS18-MS
Calibration from 20 to 4000G is possible; ideal for measuring the linearity of G; PTB traceable.
The CS18-MS is a shock-type accelerometer calibration device that uses the Hopkinson bar method. The operation is automated, providing electrical pulses to the piezoelectric element a specified number of times according to a list table, resulting in very short test times and excellent waveform reproducibility. It is capable of calibration up to a very high G level of 4,000G. Please utilize it for dynamic calibration of accelerometers and G linearity testing. This product is for secondary calibration, but we also offer primary calibration equipment.
Inquire About This Product
Related Videos
basic information
□■ Manufacturer Information ■□ SPEKTRA is a German manufacturer specializing in vibration and acoustic sensor calibration devices, producing from national primary standard calibration machines. Additionally, it is certified by DAkkS (DKD) for laboratory accreditation, allowing this product to issue PTB traceable certificates for calibration. It also complies with the internationally important standard ISO 16063-22 (secondary comparison calibration - shock). □■ Performance ■□ - Applied G: 20G to 4000G - Maximum weight of the calibrated item: 30 grams - Pulse width: 40μs to 180μs (half sine), 80μs to 360μs (full sine) - Error: 20G to 500G = less than 2.0%, up to 1000G = less than 2.5%, up to 4000G = less than 3.0% - Shunt check compatible Dynamic calibration of accelerometers and shock transducers is possible. After installation, expansion to static frequency calibration machines is also possible. Shakers, amplifiers, controllers, etc., can be customized to fit your environment. Please feel free to contact us.
Price range
Delivery Time
Applications/Examples of results
Dynamic calibration of accelerometers and shock transducers Detection of defective products that cannot be identified through static calibration and G linearity measurement Our company listens to your testing environment and requirements, and proposes the optimal system that meets your conditions. If you have any concerns about selecting a model, please feel free to contact us. Search keywords: sensor, calibration, adjustment, calibration, vibration test system, calibration machine, vibration, measurement, control, frequency, modal analysis, mount, fixture, reference meter, sine wave, G measurement, acceleration, amplitude, steady wave, B&K, Meggitt, MEGGITT, Endevco, ENDEVCO, SPEKTRA, Spectra, APS
catalog(2)
Download All CatalogsCompany information
A modest and inconspicuous thing, yet essential for existence, we tackle it with a spirit of challenge towards significant work that is rewarding, even on a small scale, while providing products that are indispensable in the future society and services that bring joy to people. As a business company that continues to develop in the 21st century, F.I.T. Pacific Co., Ltd. is a group that works tirelessly to enrich people's lives, ensure their safety, and improve the Earth's environment.