Simultaneous measurement of TOC and TN/TP at high concentration without dilution in a short time.
The BioTector series online TOC (Total Organic Carbon) analyzer can obtain accurate results even for difficult applications. It has the same functions as the B7000i in the same series, and can simultaneously measure TN/TP* without dilution, even at high concentrations. *Measurement of TN (Total Nitrogen): During the TOC analysis process, nitrogen compounds are decomposed into nitrates. The samples that have completed TOC measurement are sent for measurement, where they are analyzed using a spectrophotometric method at a wavelength of 217 nm tailored for nitrates. The flash light source of the spectrophotometer is activated, and the TN value is measured from the nitrate ions. *Measurement of TP (Total Phosphorus): During the TOC analysis process, lower phosphorus compounds are decomposed into phosphates. The samples after measurement are sent to the TP boiler, where they are heated at 100°C for about 6 minutes, allowing the remaining poly-phosphorus compounds to be heated and hydrolyzed. TP reagents are added, and the samples are further heated to 45°C before entering the measurement cell. They are measured using a spectrophotometric method at a wavelength of 405 nm tailored for phosphates. The flash light source of the spectrophotometer is activated, and the TP value is measured from the ortho-phosphate ions.
Inquire About This Product
basic information
【TOC (VOC, COD, BOD)】 ● Method: Wet two-stage oxidation method (hydroxyl radical) ● Detection method: NDIR method ● Measurement range: Depends on the combination with other measurement items. Please contact us for details. 【TN】 ● Method: Wet two-stage oxidation method (hydroxyl radical) ● Detection method: Spectrophotometry (217nm ultraviolet light) ● Measurement range: Depends on the combination with other measurement items. Please contact us for details. 【TP】 ● Method: Wet two-stage oxidation method + heating and hydrolysis ● Detection method: Spectrophotometry (405nm ultraviolet light) ● Measurement range: Depends on the combination with other measurement items. Please contact us for details. ■ Specifications such as size are common. Please download the catalog for details. ● External dimensions: W 750 × H 1,250 × D 320 (mm) *H can be up to 1,750 mm depending on options ● Weight: 90kg *Can be up to 120 kg depending on options ● External voltage: AC 100 - 120V or AC 200 - 250V 50/60Hz ● Power consumption: 300 W
Price information
Please contact us.
Delivery Time
Applications/Examples of results
**BioTector (Features of the BioTector Series)** Online TOC analyzer with unique technology and wet two-stage oxidation method Easy maintenance every six months for long-term continuous monitoring and stable operation Measurement accuracy of ±3% (reading value) Automatic range switching function and multi-channel measurement capability A variety of options available (up to 6 channels, TN/TP compatible, pretreatment devices, etc.) Calibration with standard solution twice a year Drift rate of 5% per year (On-demand seminar available. https://hach.jp/form/1289/) **BioTector: Industries with Implementation (Applications)** - Petrochemicals (high salinity, volatile organic compounds) - Oil refining (high oil content, calcium, salinity) - Machinery and metals (high turbidity) - Food and dairy products (high protein, sugar, oil content) - Pulp and paper (fibrous substances, solids) - Pharmaceuticals - Airports (chemicals, solids, salinity in de-icing liquids) - Others **Support** Technical staff with specialized knowledge provide support and solutions tailored to customer requests. - On-site training - Trial operation - Reagents (SDS) - Various information - Environment - Consulting
Detailed information
-
●Effect on running costs The introduction of a non-clogging TOC analyzer bioprotects significantly reduces the labor involved in cleaning and maintenance of various components. Due to the wet two-stage oxidation method, it lowers the total cost of TOC measurement without the need for combustion tube or catalyst replacement.
-
●Monitoring Points for TOC Online TOC has many important points that should be measured not only for wastewater management but also within the water circulation system of the factory.
-
●Sample tube diameter 3.2mm By adopting a large-diameter sample tube with an inner diameter of 3.2mm, it can accommodate particle sizes up to 2mm.
catalog(6)
Download All CatalogsCompany information
Hack Japan, as the Japanese subsidiary of Hach Company, is engaged in the sales and service provision of a portion of Hach's product lineup. Since our establishment, we have been providing high-quality analyzers that are highly accurate and easy to use. We continue to uphold the values established by founders Clifford and Kitty Hach when they founded the company in 1947. For over 60 years, we have consistently provided the best solutions for water management and have been trusted by various industries. Our analytical instruments and reagents are used for water quality testing across a range of industrial sectors. Water analysis must be accurate. Hach supports water analysis experts by providing reliable and user-friendly products. 【Main Product Brands】 - Online TOC Analyzer BioTector Series - Various Dissolved Gas Analyzers Orbisphere Series - Online VFA Analyzer EZ7200 - Silica and Sodium Analyzers Polymetron Series - Laboratory TOC Analyzer QbD1200+ - Liquid Spectrophotometer (Color Difference Meter) LICO 620/290